| Characteristics and effective method of use of Micro Surgical Blade for Fine Incision |
K-5100^K-5110 |
![]() |
| Open a small
slit in the dura mater with the blade, insert the guide part of the dura mater blade, and adjust the angle of the blade so that the tip may touch the dura mater at a right angle. |
Pinch one end
of the dura mater with a pincette, give tension to it, and cut the dura mater while pushing it. (If the cutting is difficult, readjust the angle of the blade.) |
K-5200^K-5210 |
![]() |
| Cut the arachnoid
membrane on the tissue and the blood vessel with the circular blade while gently pushing it. (If the tip of the blade cannot be seen well, use a single-side guard K-5210.) |
A safety guard is
provided, but the blade may get deeper than necessary if you push it too strongly for the cutting. Start cutting gently and then increase the cutting power little by little. |
K-5610 |
![]() |
| The arachnoid membrane running on the blood vessel can be cut safely by means of a saw-tooth tip. |
It is also possible to peel off the arachnoid membrane and the brain by using the side face of the blade. |
@
@